Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K180249

    Validate with FDA (Live)

    Date Cleared
    2018-10-11

    (255 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OmniPore® DUROMAX® Surgical Implants are intended for non-weight bearing applications of maxillofacial and orbital reconstruction, cosmetic surgery and repair of maxillofacial and orbital trauma.

    Device Description

    The OmniPore DUROMAX Surgical Implants are an extension within the family of OmniPore Surgical Implants. OmniPore DUROMAX Surgical Implants include the addition of pure titanium within the porous high-density polyethylene. They are marketed as single use sterile implants with various sizes. The material used for the OmniPore DUROMAX Surgical Implants is a porous high-density (HD) polyethylene. Polyethylene has a long history of use in surgical implantable products. The interconnecting open pore structure of the implants allow for tissue ingrowth. Both of the materials used to manufacture the OmniPore DUROMAX Surgical Implants, the porous HD polyethylene and pure titanium, have been utilized in reconstruction and soft tissue repair for many years. There is a long history of the use of porous polyethylene and titanium implants for enucleation and evisceration, as well as for reconstruction/cosmetic surgery and repair of the orbital skeletal framework, including orbital floors, walls, rims and roof, with a history of safety and performance. The interconnecting open pore structure of the OmniPore DUROMAX Surgical Implants allow for tissue ingrowth. Additionally, animal histology has shown fibrovascular ingrowth occurs into the open pore structure of porous surgical implants. The other advantage to the OmniPore DUROMAX Surgical Implants is that the titanium structure can be curved or rounded and will hold that shape in the internal orbital skeletal framework. The titanium material contained in OmniPore DUROMAX Surgical Implant is also radiopaque on post-operative CT (computed tomography) scans. The implants are single use and provided sterile by ethylene oxide (EO) terminal sterilization.

    AI/ML Overview

    Acceptance Criteria and Study for OmniPore® DUROMAX® Surgical Implants

    This document describes the acceptance criteria and the study that demonstrates the OmniPore® DUROMAX® Surgical Implants meet these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state quantitative acceptance criteria for device performance. Instead, it relies on demonstrating substantial equivalence to existing predicate devices. The performance is assessed through various tests aligning with recognized standards and a comparison of technological characteristics.

    Acceptance Criteria CategorySpecific Criteria (Implicit via Substantial Equivalence)Reported Device Performance
    BiocompatibilityCompliance with ISO 10993-1, 5, 10, 11Demonstrated compliance through completed Cytotoxicity, Irritation, and Material-Mediated Pyrogenicity testing. Justification for HDPE (per K123908) and pure titanium (per ASTM F67 and ISO 5832-2) maintaining compliance through manufacturing and sterilization.
    SterilizationCompliance with ISO 11135 (SAL at 10⁻⁶) & ISO 10993-7 (EO Residuals)Justified compliance against previous sterilization validation per ISO 11135 (SAL at 10⁻⁶) and EO Residuals requirements per ISO 10993-7.
    PyrogenicityNon-pyrogenic claimDemonstrated non-pyrogenic per USP <161>, USP <85>, and USP <151>.
    Shelf Life5-year shelf lifeJustified compliance against previous five (5) year shelf life testing.
    Mechanical PropertiesNot explicitly defined, but inferred from predicate devices and specific tests.Impact testing per ASTM D5420 (justified against previous testing), Purity testing per USP <661>, Porosity and Ingrowth testing per ASTM D4284. Additional bench testing: bend, trim/cut, globe support, and fixation.
    Material CompositionPorous high-density polyethylene and pure titaniumConfirmed use of porous high-density (HD) polyethylene and pure titanium.
    RadiopacityRadiopaque on post-operative CT scansConfirmed radiopaque property of titanium material on post-operative CT scans.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes material and mechanical testing, not a clinical study with a "test set" in the sense of patient data.

    • Sample Size for Material/Mechanical Testing: Not explicitly stated for each test (e.g., how many implants were subjected to bend testing). The document states "Compliance was justified against previous five (5) year shelf life testing and previous mechanical testing," implying these tests were performed on an adequate number of samples to represent the device.
    • Data Provenance: The data provenance is from laboratory bench testing and refers to recognized international and US standards (ISO, ASTM, USP). This is not patient data, so concepts like country of origin or retrospective/prospective do not apply.

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    Not applicable as this is not a study involving expert interpretation of medical images or patient outcomes. The "ground truth" for material and mechanical properties is established by the specifications defined in the referenced standards (e.g., ISO 10993, ASTM F67).

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used in clinical studies for resolving discrepancies in expert interpretations of medical data. For material and mechanical testing, the "ground truth" is determined by adherence to specified test methodologies and resulting measurements against predefined acceptance criteria from standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not a diagnostic device or an AI-assisted device requiring human-in-the-loop performance evaluation. The device is a surgical implant.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This is a physical surgical implant, not an algorithm or AI system.

    7. Type of Ground Truth Used (for all studies)

    The ground truth for the safety and performance evaluation of the OmniPore® DUROMAX® Surgical Implants is based on:

    • International and US Standards: Such as ISO 10993 series (biocompatibility), ISO 11135 (sterilization), ISO 10993-7 (EO residuals), USP <161>, <85>, <151> (pyrogenicity), ASTM D5420 (impact testing), USP <661> (purity), and ASTM D4284 (porosity and ingrowth).
    • Established Performance of Predicate Devices: The concept of substantial equivalence heavily relies on demonstrating that the new device shares similar technological characteristics and has similar indications for use as legally marketed predicate devices with a known history of safety and effectiveness.

    8. Sample Size for the Training Set

    Not applicable. There is no "training set" as this device is a physical implant evaluated through material characterization and mechanical testing, not an AI or machine learning model.

    9. How the Ground Truth for the Training Set was Established

    Not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133046

    Validate with FDA (Live)

    Date Cleared
    2014-05-14

    (229 days)

    Product Code
    Regulation Number
    878.3500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OmniPore® Customized Surgical Implants are intended for non-weight bearing applications of craniofacial reconstruction/ cosmetic surgery and repair of craniofacial trauma. OmniPore Customized Surgical Implants are also intended for the augmentation or restoration of contour in the craniomaxillofacial skeleton.

    Device Description

    The OmniPore® Customized Surgical Implants are marketed as single patient use sterile implants that physicians request as customized surgical implants which use identical materials and manufacturing to the OmniPore® Surgical Implants but are made to a predetermined patient's measurements and size requirements. The applications include non-load bearing augmentation and/or reconstruction of the craniofacial skeleton. The OmniPore Customized Surgical Implants are created from the patient's CT imaging data provided from the physician. The OmniPore Customized Surgical Implants are manufactured from the same material, manufactured under the same processes, and packaged the same as the OmniPore Surgical Implants.

    The raw material used for the OmniPore Customized Surgical Implants is high-density polyethylene resin which has a long history of use in surgical implantable products. The interconnecting open pore structure of the OmniPore Customized Surgical Implants allow for tissue in-growth. Additionally, animal histology has shown fibrovascular in-growth occurs into the open pore structure of OmniPore Customized Surgical Implants.

    The implants are single use and provided sterile by ethylene oxide (EO) terminal sterilization.

    AI/ML Overview

    This document describes the OmniPore® Customized Surgical Implants, a device intended for non-weight-bearing applications in craniofacial reconstruction/cosmetic surgery, repair of craniofacial trauma, and augmentation/restoration of contour in the craniomaxillofacial skeleton.

    Here's an analysis of the provided information regarding acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state specific quantifiable acceptance criteria for the device's performance. Instead, it relies on demonstrating substantial equivalence to predicate devices. The "reported device performance" is implicitly that the device performs equivalently to its predicates across various aspects.

    Acceptance Criteria CategoryAcceptance Criteria (Explicitly Stated in Document)Reported Device Performance
    Material EquivalenceNot explicitly stated as a quantitative criterion, but implicit expectation is the materials are identical or perform equivalently to predicate devices."The OmniPore® Customized Surgical Implant materials are equivalent materials as the previously cleared OmniPore Surgical Implant devices..."
    BiocompatibilityNot explicitly stated as a quantitative criterion, but implicit expectation is the device is biocompatible as demonstrated by predicate testing."...so the biocompatibility and sterilization validation to validate that they are sterile devices for implantation was justified from the predicate testing." (Performance is inferred to be adequate based on predicate.)
    Sterilization EfficacyNot explicitly stated as a quantitative criterion, but implicit expectation is terminal sterilization is effective (e.g., Sterility Assurance Level of 10^-6)."...so the biocompatibility and sterilization validation to validate that they are sterile devices for implantation was justified from the predicate testing." (Performance is inferred to be adequate based on predicate.) The implants are "provided sterile by ethylene oxide (EO) terminal sterilization."
    Mechanical Properties (Impact, Purity, Porosity)Not explicitly stated as a quantitative criterion, but implicit expectation is that these properties are equivalent to predicate devices."The OmniPore Customized Surgical Implants are equivalent materials as the previously cleared OmniPore Surgical Implant devices so the mechanical testing specific to impact testing, purity testing per USP, and porosity testing was justified from the OmniPore Surgical Implant device testing." (Performance is inferred to be adequate based on predicate.)
    Software Verification (for Customization Process)Not explicitly stated, but implicit expectation is that the software used for customized design is validated and functions equivalently to predicate."The software verification testing documentation for the OmniPore Customized Surgical Implants was provided to justify equivalence against the MEDPOR® Customized Surgical Implant." (Performance is inferred to be adequate based on predicate, indicating the software part of the device's manufacturing process has been verified.)
    Clinical Performance / Safety and Effectiveness for Intended Use (Overall Substantial Equivalence)No specific quantitative clinical endpoints or safety profiles are provided as acceptance criteria in this document."The OmniPore® Customized Surgical Implants have the same intended use and indications for use, and the same technological characteristics and principles of operation as the predicate devices. The minor differences do not raise any issues of safety or effectiveness. Testing results support the determination of substantial equivalence with the results demonstrating that the OmniPore Customized Surgical Implants have equivalent results as the predicate devices." (Summary conclusion of equivalence.)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a "test set" in the traditional sense of a clinical or retrospective study with patient data for evaluating the device's performance. The evaluation for this 510(k) submission is based on demonstrating equivalence to predicate devices, primarily through material and manufacturing equivalency, and justification from existing predicate testing data.

    Therefore, information on sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable or provided in this type of submission which focuses on substantial equivalence rather than a de novo clinical study. The "testing" mentioned refers to justification based on previously conducted tests for predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as there is no mention of a clinical "test set" requiring expert ground truth establishment for patient outcomes or diagnoses. The submission relies on scientific and engineering evaluations for material equivalence and manufacturing processes.

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no mention of a "test set" requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable. The device described is a physical surgical implant, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is irrelevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. The device is a physical surgical implant, not an algorithm.

    7. The Type of Ground Truth Used

    Given the nature of the submission (demonstrating substantial equivalence for a physical implant), the "ground truth" is largely:

    • Material properties and manufacturing process specifications: Confirming the raw materials, manufacturing methods, and sterilization processes are the same or equivalent to the predicate devices.
    • Existing predicate device performance data: Relying on the established safety and effectiveness of the legally marketed predicate devices through their prior biocompatibility, sterilization validation, and mechanical testing results.
    • Software verification documentation: For the customization process, ensuring the software performs its intended function for creating the customized implant based on CT data, likely verified against functional requirements and possibly accuracy checks against reference CAD models or similar.

    There is no mention of pathology, expert consensus on patient outcomes from a new study, or specific outcomes data for the OmniPore® Customized Surgical Implants itself beyond the predicate justification.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is not an AI/ML algorithm that requires a training set. The customization is based on patient-specific CT imaging data for manufacturing, not on a machine learning model trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1